Publication:
Mycobacterium chimaera in heater-cooler devices: an experience in a tertiary hospital in Spain

dc.contributor.authorQuintás Viqueira, A
dc.contributor.authorPérez Romero, César
dc.contributor.authorToro Rueda, C
dc.contributor.authorSánchez Calles, A M
dc.contributor.authorBlázquez González, J A
dc.contributor.authorAlejandre Leyva, M
dc.date.accessioned2021-01-19T09:18:51Z
dc.date.available2021-01-19T09:18:51Z
dc.date.issued2021-01
dc.description.abstractThe aim of the study was to describe the Mycobacterium chimaera contamination in heater-cooler devices after the application of a protocol of cleaning and disinfection in a tertiary hospital. It was an observational study at the La Paz-Cantoblanco-Carlos III University Hospital, Madrid, Spain. Seven heater-cooler devices are used in our hospital: five 3T Sorin (LivaNova) and two Maquet. We followed the manufacturers' instructions for cleaning and disinfection of the different heater-cooler devices. Environmental testing was developed monthly from January 2017 to July 2019. Samples were obtained from both cardioplegia and patient circuits and before and after the disinfection process and were cultured in appropriate media for non-tuberculous mycobacteria and heterotrophic bacteria (coliforms and Pseudomonas aeruginosa). A total of 320 samples were taken. Mycobacterium chimaera grew in four water samples (1.25%) from three different heater-cooler devices, with two positive results occurring after disinfection. The heterotrophic bacteria Delftia acidovorans and Stenotrophomonas maltophilia were also found. There has not been a case of M. chimaera infection in patients after cardiac surgery in our hospital. In March 2019, we decided to move the heater-cooler device outside the operating room. Mycobacterium chimaera contamination is not always eradicated by disinfection processes. We believe that placing 3T heater-cooler devices outside the operating room is the best option in preventing M. chimaera infection during cardiac surgery.es_ES
dc.description.peerreviewedes_ES
dc.format.page100757es_ES
dc.format.volume39es_ES
dc.identifier.citationNew Microbes New Infect . 2020 Sep 7;39:100757.es_ES
dc.identifier.doi10.1016/j.nmni.2020.100757es_ES
dc.identifier.e-issn2052-2975
dc.identifier.journalNew microbes and new infectionses_ES
dc.identifier.pubmedID33437492es_ES
dc.identifier.urihttp://hdl.handle.net/20.500.12105/11634
dc.language.isoenges_ES
dc.publisherElsevier
dc.relation.publisherversionhttps://doi.org/10.1016/j.nmni.2020.100757es_ES
dc.repisalud.centroISCIII::Escuela Nacional de Sanidad (ENS)es_ES
dc.repisalud.institucionISCIIIes_ES
dc.rights.accessRightsopen accesses_ES
dc.rights.licenseAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subjectAerosolizationes_ES
dc.subjectCardiac surgeryes_ES
dc.subjectDisinfectiones_ES
dc.subjectHeater-cooler devicees_ES
dc.subjectMycobacterium chimaeraes_ES
dc.titleMycobacterium chimaera in heater-cooler devices: an experience in a tertiary hospital in Spaines_ES
dc.typeresearch articlees_ES
dc.type.hasVersionVoRes_ES
dspace.entity.typePublication
relation.isAuthorOfPublication3435f2c5-4f0a-4c65-86f0-c070ebd5fb27
relation.isAuthorOfPublication.latestForDiscovery3435f2c5-4f0a-4c65-86f0-c070ebd5fb27
relation.isPublisherOfPublication7d471502-7bd5-4f7a-90a4-8274382509ef
relation.isPublisherOfPublication.latestForDiscovery7d471502-7bd5-4f7a-90a4-8274382509ef

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
MycobacteriumChimaeraInHeater_2021.pdf
Size:
651.61 KB
Format:
Adobe Portable Document Format
Description: